title: "Sample Holding Times: How a LIMS Tracks Them - Prisma LIMS"
description: "Sample holding times in plain words: when the clock starts, common EPA hold times with sources, prep and analysis clocks, and how a LIMS tracks them."
image: https://prismalims.com/hubfs/website%20assets/images/featured%20images/og-sample-holding-times.jpg
---

[Skip to content](https://prismalims.com/sample-holding-times#main)

[![Prisma LIMS](https://prismalims.com/hs-fs/hubfs/website%20assets/logos/prisma-lims-logo-1-line-340.webp?width=170&height=30&name=prisma-lims-logo-1-line-340.webp)](https://prismalims.com/) [Schedule an online demo](https://meetings.hubspot.com/dacian-florea)

- [Home](https://prismalims.com/)
- [How You Get Live](https://prismalims.com/how-it-works)
- [Who It’s For](https://prismalims.com/who-its-for)
- Labs we serve
  
    - [Environmental LIMS](https://prismalims.com/environmental-lims)
    - By sample type 
          - [Water LIMS](https://prismalims.com/who-its-for#water)
          - [Wastewater LIMS](https://prismalims.com/wastewater-lims)
          - [Drinking water LIMS](https://prismalims.com/drinking-water-lims)
          - [Soil, sludge and biosolids](https://prismalims.com/soil-lims)
    - By lab type 
          - [Commercial environmental labs](https://prismalims.com/commercial-lab-lims)
          - [Municipal and treatment-plant labs](https://prismalims.com/municipal-lab-lims)
- Guides
  
    - [Cloud LIMS](https://prismalims.com/cloud-lims)
    - [Chain of custody software](https://prismalims.com/chain-of-custody-software)
    - [Sample holding times](https://prismalims.com/sample-holding-times)
    - [TCLP and SPLP](https://prismalims.com/tclp-splp-leaching)
    - [PFAS testing](https://prismalims.com/pfas-testing-lims)
    - [NELAP and TNI records](https://prismalims.com/nelap-tni-lims-requirements)
    - [LIMS cost over five years](https://prismalims.com/lims-total-cost-of-ownership)
    - [Compare environmental LIMS vendors](https://prismalims.com/compare)
    - [Blog](https://prismalims.com/blog)
- [About Us](https://prismalims.com/about)

[Schedule an online demo](https://meetings.hubspot.com/dacian-florea)

1. [Home](https://prismalims.com/)
2. [Environmental LIMS](https://prismalims.com/environmental-lims)
3. Sample holding times

Clock, batch, QC

# Sample holding times: how a LIMS keeps the clock

A sample holding time is the longest time allowed between collecting a sample and preparing or analyzing it. At 6 a.m. on Wednesday, an analyst finds two BOD bottles from Monday’s 7 a.m. sampling at the back of the walk-in cooler. The 48-hour clock started at collection (40 CFR 136.3, Table II), so there is one hour left. This guide is for lab managers, QA staff and analysts in water, wastewater and soil labs. When does the clock start, how long is it, and what should your lab software do about it?

By the Prisma LIMS team · Updated 26 November 2026

[Schedule an online demo](https://meetings.hubspot.com/dacian-florea) [See the hold time table](https://prismalims.com/sample-holding-times#hold-times)

In short

A sample holding time is the longest time allowed between collecting a sample and preparing or analyzing it. The clock starts at collection, and some tests have two clocks: one to extraction and one to analysis. A LIMS keeps each clock from the collection time on the chain of custody and shows what is about to run out. Prisma LIMS supports hold-time tracking, batching and QC, and we set them up for your methods.

On this page

1. [Why one missed hold time costs a day](https://prismalims.com/sample-holding-times#why)
2. [What a holding time is](https://prismalims.com/sample-holding-times#what)
3. [Common EPA holding times](https://prismalims.com/sample-holding-times#hold-times)
4. [How a LIMS tracks holding times](https://prismalims.com/sample-holding-times#how)
5. [Batches, QC and the MDL](https://prismalims.com/sample-holding-times#batch-qc)
6. [What to test in a demo](https://prismalims.com/sample-holding-times#demo)
7. [How Prisma LIMS supports it](https://prismalims.com/sample-holding-times#prisma)
8. [Questions people ask](https://prismalims.com/sample-holding-times#questions)

Why it matters

## Why does one missed hold time cost a day?

A sample past its hold time is a lost result. Someone has to call the client. The client has to send a crew back out to sample again.

For a coliform sample under the Clean Water Act, the clock is 8 hours from collection (40 CFR 136.3, Table II). It can run out before the sample even reaches the bench. One missed clock costs a day of field work, a resample, and a bit of the client’s trust.

Working harder does not fix it. Seeing every clock in one place, early enough to act, does. That is the job of your LIMS.

The basics

## What is a sample holding time?

A holding time is the longest a sample may wait before the lab starts the test and still count as valid. Under the Clean Water Act, the times in the rule are “the maximum times that samples may be held before the start of analysis and still be considered valid” (40 CFR 136.3, Table II).

The same rule says the table “takes precedence over information in specific methods” (40 CFR 136.3(e)). So for permit work, the table wins over the method.

### When does the clock start?

The Clean Water Act table sets the start point by sample type (40 CFR 136.3, Table II, footnote 4):

- For a grab sample, the clock starts at the time of collection.
- For a composite from an automatic sampler, it starts at the end of collection.
- For grab samples composited later, it starts when the last grab in the set is collected.

So every clock starts from the [chain of custody record](https://prismalims.com/chain-of-custody-software).

### Why do some tests have two clocks?

Many organic tests have two clocks. The first runs from collection to extraction. The second runs from extraction to analysis of the extract. The same idea holds for leaching tests such as [TCLP and SPLP, which add an extraction step](https://prismalims.com/tclp-splp-leaching).

Clock one: collection to extraction

Clock two: extraction to analysis

Example from the Clean Water Act table for pesticides, PAHs and phthalate esters in water. Extract within 7 days of collection. Then analyze the extract within 40 days (40 CFR 136.3, Table II).

Look it up

## What are the EPA holding times for common tests?

The limit depends on the program your sample is for. A coliform sample for a wastewater permit and one for a drinking water system follow different rules.

| Test | Program | Maximum holding time | Source |
| --- | --- | --- | --- |
| Coliform (total, fecal) and E. coli | Clean Water Act (40 CFR 136.3) | 8 hours. Incubation must start no later than 8 hours from collection. | 40 CFR 136.3, Table II |
| Biochemical oxygen demand (BOD and carbonaceous BOD) | Clean Water Act (40 CFR 136.3) | 48 hours | 40 CFR 136.3, Table II |
| Nitrate, or nitrite | Clean Water Act (40 CFR 136.3) | 48 hours each. Nitrate-nitrite preserved with acid: 28 days. | 40 CFR 136.3, Table II |
| Orthophosphate | Clean Water Act (40 CFR 136.3) | Filter within 15 minutes. Analyze within 48 hours. | 40 CFR 136.3, Table II |
| pH, total residual chlorine, dissolved oxygen (probe), temperature | Clean Water Act (40 CFR 136.3) | Analyze within 15 minutes. | 40 CFR 136.3, Table II |
| Total suspended solids (TSS) | Clean Water Act (40 CFR 136.3) | 7 days | 40 CFR 136.3, Table II |
| Ammonia; total phosphorus | Clean Water Act (40 CFR 136.3) | 28 days (preserved with acid) | 40 CFR 136.3, Table II |
| Cyanide | Clean Water Act (40 CFR 136.3) | 14 days | 40 CFR 136.3, Table II |
| Metals, except boron, chromium VI and mercury | Clean Water Act (40 CFR 136.3) | 6 months | 40 CFR 136.3, Table II |
| Mercury | Clean Water Act (40 CFR 136.3) | 28 days (cold vapor AA); 90 days (cold vapor AFS) | 40 CFR 136.3, Table II |
| Chromium VI | Clean Water Act (40 CFR 136.3) | 28 days, with the buffer the table names | 40 CFR 136.3, Table II |
| Pesticides, PAHs, phthalate esters (water) | Clean Water Act (40 CFR 136.3) | 7 days until extraction, 40 days after extraction | 40 CFR 136.3, Table II |
| Purgeable aromatics, such as benzene (water) | Clean Water Act (40 CFR 136.3) | 14 days with acid to pH 2. 7 days without it. | 40 CFR 136.3, Table II |
| Total coliform and E. coli, drinking water samples | Safe Drinking Water Act (40 CFR 141) | 30 hours from collection to the start of incubation | 40 CFR 141.852(a)(3) |
| Nitrate, drinking water samples | Safe Drinking Water Act (40 CFR 141) | 48 hours. 14 days if the sample is chlorinated, not acidified, and kept at 4 °C. | 40 CFR 141.23(k)(2) |
| Volatile organics in water, acid to pH below 2 | SW-846 (EPA guidance) | 14 days | SW-846 Chapter Four, Table 4-1 |
| Semivolatile organics in water | SW-846 (EPA guidance) | Extract within 7 days. Analyze the extract within 40 days. | SW-846 Chapter Four, Table 4-1 |
| Semivolatile organics in soil, sediment or sludge | SW-846 (EPA guidance) | Extract within 14 days. Analyze the extract within 40 days. | SW-846 Chapter Four, Table 4-1 |

Your permit, your state or your method may set a shorter time. Always check the current rule.

PFAS samples follow the hold times in each PFAS method. See [PFAS testing and Methods 1633, 533 and 537.1](https://prismalims.com/pfas-testing-lims).

The table also lists the container and the preservative for each test. A sample that was not cooled or preserved the right way can fail even inside its time.

SW-846 times are guidance, not a rule. EPA says Table 4-1 “does not represent EPA requirements” (SW-846 Chapter Four). Your project plan or your state sets which times apply.

Keep the clock

## How does a LIMS track sample holding times?

Any LIMS for an environmental lab should do these things for hold times.

- The clock starts from the date and time on the chain of custody, not from the time of login.
- The system knows the sample type. A composite starts at the end of collection, so it needs both times.
- Each test and program carries its own limit, set once per method.
- Tests that need it run two clocks: one to extraction, and one from extraction to analysis.
- One screen shows every open sample, with the time left on each clock.
- The shortest clock goes to the top of the bench’s work list.
- A warning goes out while there is still time to run the test.
- If a test starts past its limit, the report says so.

Ask any vendor

- Does the clock start from the collection time on the COC, or from login?
- Who sets the hold time for each method, and is each change on the audit trail?
- Then run the [demo list below](https://prismalims.com/sample-holding-times#demo): composites, two clocks, time left and late results.

Batch and QC

## Batches and QC: what rides with the clock

The clock decides when a sample must go into a batch. The batch decides which QC it gets. Labs we talk to often say QC review takes a big share of the day.

### What is batch QC? Blanks, LCS, MS/MSD and duplicates

Batch QC is the set of checks that runs with each group of samples. It shows the batch can be trusted. For Clean Water Act methods that lack their own QC steps, EPA lists twelve QC elements a lab must cover (40 CFR 136.7). The ones you see in every batch:

- A method blank is clean water taken through every step. It shows the lab did not add the analyte.
- A laboratory control sample (LCS) is a clean sample spiked with a known amount. It shows the method works.
- A matrix spike and matrix spike duplicate (MS/MSD) are a real sample, spiked twice. They show how the sample itself affects the result.
- A duplicate is the same sample run twice. It shows how close repeat results are.

### Prep batches and analytical runs: what is the difference?

A prep batch is the group of samples prepared together, for example extracted or digested at the same time. An analytical run is what goes on the instrument. One prep batch can span two runs, and one run can hold several prep batches.

EPA names “definitions of preparation and analytical batches that may drive QC frequencies” as a QC element (40 CFR 136.7). A LIMS must link each sample to both its prep batch and its run. Then the right QC follows each sample.

### Who sets the control limits?

Each QC check has control limits: the range its result must fall in. Someone enters those limits for each method at set-up. Later, someone may need to change them. A LIMS should record who did it and when, on the audit trail.

### What is the method detection limit (MDL)?

The MDL is “the minimum measured concentration of a substance that can be reported with 99% confidence that the measured concentration is distinguishable from method blank results” (40 CFR 136, Appendix B).

To set a first MDL, a lab runs at least seven spiked samples and seven method blanks. They must be prepared in at least three batches, on three separate days. The lab then checks the MDL at least once every thirteen months (Appendix B).

So a LIMS should keep an MDL for each method, analyte and matrix. It should also keep the spike and blank results the MDL came from.

### J values: results between the MDL and the reporting limit

A result above the MDL but below the reporting limit is a real detection, but its number is less sure. Many labs report it with a J flag. In EPA’s data review guidelines, J means “the result is an estimated quantity”. The LIMS should add the flag on its own, from the MDL and reporting limit it holds.

### Extra blanks and one run for several methods

After a very dirty sample, such as landfill leachate, an analyst may run extra blanks to clear the instrument. The run must accept a blank anywhere, not only in fixed spots.

One run can also serve more than one method. The LIMS must still judge the QC for each method on its own. Our [guide to leaching tests and metals](https://prismalims.com/tclp-splp-leaching) shows this for metals work.

Ask any vendor

- Who sets our control limits, and can our admin change them?
- Show us an MDL kept per method, analyte and matrix.
- Can an analyst add extra blanks anywhere in a run?
- Show us one run that serves two methods, with QC judged per method.

Bring this list

## What should you test in a LIMS demo?

Print this list. Ask every vendor to do each step on your own methods, not on a demo method.

- Start with your shortest clock: log a coliform sample with its collection time. Where does the clock show?
- Enter the start and end of collection for a composite. Which one starts the clock?
- Log a sample that needs extraction. Are both clocks shown?
- Open the bench view. Is it sorted by time left?
- Ask when the warning goes out, and to whom.
- Build a batch. Do the blank, LCS, MS/MSD and duplicate come with it?
- Show a failed QC check. What happens to the samples in that batch?
- Report a low result, between the MDL and the reporting limit. Is the J flag added for you?
- Change a control limit. Is the change on the audit trail?
- Start a test after its hold time. What does the report say?

![Prisma LIMS incubator screen: coliform samples with incubation timers and status, from incubating to max incubation done](https://prismalims.com/hs-fs/hubfs/website%20assets/images/features-operational-control-800.webp?width=2000&height=1397&name=features-operational-control-800.webp)

The Prisma LIMS incubator view: each coliform sample with its timer and status.

Our part

## How Prisma LIMS supports hold-time tracking

Prisma LIMS tracks each sample’s hold time from the collection time, shows the time left and warns you before it runs out.

One sample can carry two clocks: one to extraction and one to analysis.

Prisma LIMS runs analytical and prep batches with their QC, with warnings on both.

We set control limits at setup with your analyst. Your lab admin can change them later.

We map your workflow, set up your methods and reports, move your data, test it with you in a sandbox, train your team on-site, and stay with you through go-live.

Accreditation belongs to the lab. Prisma LIMS keeps chain of custody and a full audit trail of who changed what and when.

Ask us to show it on your own methods and files.

Back to the bench

## Fewer resamples, fewer surprises

When every clock is on one screen, the bench works the shortest clock first. QC rides with each batch, so a failed check shows before the report goes out. That means fewer resamples and fewer surprise calls to clients.

### Built for

Prisma LIMS is built only for environmental labs: mid-sized and large commercial, municipal and utility labs that test water, wastewater, drinking water, soil and sludge, in the US and Canada. Other environmental matrices are on request.

Methods running on Prisma LIMS today: EPA 524, 525 and 552, and THMs; EPA 200.7 and 200.8, and SW-846 6010 and 6020; total coliform and E. coli; BOD and TSS.

[See who Prisma LIMS fits](https://prismalims.com/who-its-for)

### Not a fit for

Prisma LIMS is not a fit for clinical-only or pharma-only labs, very small labs, or labs that prefer to build and run their own system.

[Schedule an online demo](https://meetings.hubspot.com/dacian-florea)

## Questions people ask

When does a sample’s hold time start?

For a grab sample, the hold time starts at the time of collection. For a composite from an automatic sampler, it starts at the end of collection. For grab samples composited later, it starts when the last grab is collected. That is the Clean Water Act rule (40 CFR 136.3, Table II, footnote 4). For the reasons behind it, read [why a hold time starts at collection](https://prismalims.com/blog/why-a-hold-time-starts-at-collection).

What happens if a sample is past its hold time?

Under the Clean Water Act, a sample tested after its hold time is not considered valid (40 CFR 136.3, Table II, footnote 4). Most often the client has to take a new sample. If a late result is reported at all, the report should say it was late. Your permit writer, state or client decides what they accept.

How do I calculate the method detection limit?

Run at least seven spiked samples and seven method blanks, prepared in at least three batches on three separate days. Work out the MDL from the spikes (the standard deviation times a Student’s t value) and from the blanks. The MDL is the higher of the two (40 CFR 136, Appendix B).

What is the EPA’s method detection limit (MDL) rule?

It is 40 CFR part 136, Appendix B, Revision 2. It defines the MDL and sets how a lab finds it and keeps it current. The lab collects spiked samples each quarter it runs samples and checks the MDL at least once every thirteen months (Appendix B).

What is batch QC?

Batch QC is the set of checks run with each group of samples. The usual set is a method blank, a laboratory control sample, a matrix spike and matrix spike duplicate, and a duplicate. EPA lists these among the QC elements a lab must cover when a method lacks its own (40 CFR 136.7). The batch passes only when its checks fall inside their control limits.

Is the coliform hold time the same for wastewater and drinking water?

No. For Clean Water Act samples, incubation must start within 8 hours of collection (40 CFR 136.3). Drinking water samples follow the Revised Total Coliform Rule. There, the limit is 30 hours from collection to the start of incubation (40 CFR 141.852).

## Keep reading

- ### TCLP and SPLP
  
  Leaching tests add an extraction step. That means a parent sample, a leachate and two clocks to track.
  
  [TCLP, SPLP and their two clocks](https://prismalims.com/tclp-splp-leaching)
- ### Wastewater LIMS
  
  Short bacteria and BOD clocks meet monthly permit reports. See what a wastewater lab needs.
  
  [Wastewater lab software, explained](https://prismalims.com/wastewater-lims)
- ### Drinking water LIMS
  
  A different coliform clock, plus state files and reports. See what a drinking water lab needs.
  
  [Drinking water lab software](https://prismalims.com/drinking-water-lims)
- ### Chain of custody
  
  Every clock starts from the collection time on the COC. See what a COC must record.
  
  [Environmental chain of custody software](https://prismalims.com/chain-of-custody-software)
- ### Environmental LIMS guide
  
  Hold times are one part of the job. See what else an environmental LIMS must handle.
  
  [How to compare environmental LIMS](https://prismalims.com/environmental-lims)
- ### Why a hold time starts at collection
  
  Time in the cooler and on the truck counts. A short read on the rule and why it matters.
  
  [Why a hold time starts at collection](https://prismalims.com/blog/why-a-hold-time-starts-at-collection)
- ### NELAP and TNI records
  
  Assessors check that each result met its hold time. See the records a TNI or NELAP lab must keep.
  
  [NELAP and TNI requirements for your LIMS](https://prismalims.com/nelap-tni-lims-requirements)

Last checked: 24 November 2026. Next check due: 24 May 2027.

## Schedule an online demo

The first call is a 30-minute Lab Systems Diagnostic. Then we show Prisma LIMS on your own files.

[Schedule an online demo](https://meetings.hubspot.com/dacian-florea)

[contact@prismalims.com](mailto:contact@prismalims.com) [(224) 347-6794](tel:+12243476794)

[![](https://prismalims.com/hs-fs/hubfs/website%20assets/logos/prisma-lims-logo-1-line-white-340.webp?width=170&height=30&name=prisma-lims-logo-1-line-white-340.webp)](https://prismalims.com/)

A cloud LIMS built only for environmental labs in the US and Canada. Based in Illinois, US.

Book a call

[Schedule an online demo](https://meetings.hubspot.com/dacian-florea)

Email or call

[contact@prismalims.com](mailto:contact@prismalims.com)

[(224) 347-6794](tel:+12243476794)

The first call is a 30-minute Lab Systems Diagnostic.

Company

- [Home](https://prismalims.com/)
- [How You Get Live](https://prismalims.com/how-it-works)
- [Pricing](https://prismalims.com/pricing)
- [Who It’s For](https://prismalims.com/who-its-for)
- [About Us](https://prismalims.com/about)
- [Blog](https://prismalims.com/blog)

Labs we serve

- [Environmental LIMS](https://prismalims.com/environmental-lims)
- [Commercial environmental labs](https://prismalims.com/commercial-lab-lims)
- [Municipal and treatment-plant labs](https://prismalims.com/municipal-lab-lims)

Sample types

- [Water LIMS](https://prismalims.com/who-its-for#water)
- [Wastewater LIMS](https://prismalims.com/wastewater-lims)
- [Drinking water LIMS](https://prismalims.com/drinking-water-lims)
- [Soil, sludge and biosolids](https://prismalims.com/soil-lims)

Guides

- [Cloud LIMS](https://prismalims.com/cloud-lims)
- [Chain of custody software](https://prismalims.com/chain-of-custody-software)
- [Sample holding times](https://prismalims.com/sample-holding-times)
- [TCLP and SPLP](https://prismalims.com/tclp-splp-leaching)
- [PFAS testing](https://prismalims.com/pfas-testing-lims)
- [NELAP and TNI records](https://prismalims.com/nelap-tni-lims-requirements)
- [LIMS cost over five years](https://prismalims.com/lims-total-cost-of-ownership)
- [Compare environmental LIMS vendors](https://prismalims.com/compare)

© 2026 Prisma LIMS. [Cookie Policy](https://prismalims.com/cookie-policy)

- [![](https://prismalims.com/hs-fs/hubfs/website%20assets/icons/linkedin-48.webp?width=24&height=24&name=linkedin-48.webp)](https://www.linkedin.com/company/prismalims/)
- [![](https://prismalims.com/hs-fs/hubfs/website%20assets/icons/youtube-48.webp?width=24&height=24&name=youtube-48.webp)](https://www.youtube.com/@PrismaLIMS)
- [![](https://prismalims.com/hs-fs/hubfs/website%20assets/icons/facebook-48.webp?width=24&height=24&name=facebook-48.webp)](https://www.facebook.com/prismalims/)
- [![](https://prismalims.com/hs-fs/hubfs/website%20assets/icons/instagram-48.webp?width=24&height=24&name=instagram-48.webp)](https://www.instagram.com/prismalims/)

```json
{
  "@context" : "https://schema.org",
  "@id" : "https://prismalims.com/#organization",
  "@type" : "Organization",
  "address" : {
    "@type" : "PostalAddress",
    "addressCountry" : "US",
    "addressRegion" : "IL"
  },
  "alternateName" : [ "PrismaLIMS", "LIMs-Plus" ],
  "areaServed" : [ {
    "@type" : "Country",
    "name" : "United States"
  }, {
    "@type" : "Country",
    "name" : "Canada"
  } ],
  "contactPoint" : {
    "@type" : "ContactPoint",
    "availableLanguage" : "English",
    "contactType" : "sales",
    "email" : "contact@prismalims.com",
    "telephone" : "+1-224-347-6794"
  },
  "description" : "Prisma LIMS is a cloud LIMS for environmental testing labs in the US and Canada. We set it up for your lab, move your data, train your team, and run the hosting and support.",
  "email" : "contact@prismalims.com",
  "founder" : {
    "@id" : "https://prismalims.com/about#dacian-florea",
    "@type" : "Person",
    "name" : "Dacian Florea",
    "sameAs" : [ "https://www.linkedin.com/in/dacian-florea/" ]
  },
  "foundingDate" : "2021",
  "knowsAbout" : [ "Laboratory information management systems", "Environmental laboratory software", "Cloud LIMS", "Wastewater testing", "Drinking water testing", "Soil and sludge testing", "Chain of custody", "Sample holding times", "TCLP and SPLP leaching", "PFAS testing methods", "TNI and NELAP accreditation records" ],
  "logo" : {
    "@id" : "https://prismalims.com/#logo",
    "@type" : "ImageObject",
    "caption" : "Prisma LIMS",
    "height" : 512,
    "url" : "https://48211686.fs1.hubspotusercontent-na1.net/hubfs/48211686/website%20assets/logos/prisma-lims-favicon-512.png",
    "width" : 512
  },
  "name" : "Prisma LIMS",
  "sameAs" : [ "https://www.linkedin.com/company/prismalims/", "https://www.youtube.com/@PrismaLIMS", "https://www.facebook.com/prismalims/", "https://www.instagram.com/prismalims/" ],
  "telephone" : "+1-224-347-6794",
  "url" : "https://prismalims.com/"
}
```

```json
{
  "@context" : "https://schema.org",
  "@id" : "https://prismalims.com/#website",
  "@type" : "WebSite",
  "alternateName" : [ "PrismaLIMS", "LIMs-Plus" ],
  "inLanguage" : "en-US",
  "name" : "Prisma LIMS",
  "publisher" : {
    "@id" : "https://prismalims.com/#organization"
  },
  "url" : "https://prismalims.com/"
}
```

```json
{
  "@context" : "https://schema.org",
  "@id" : "https://prismalims.com/sample-holding-times#webpage",
  "@type" : "WebPage",
  "about" : [ {
    "@id" : "https://prismalims.com/#organization"
  }, {
    "@id" : "https://prismalims.com/sample-holding-times#term-holding-time"
  } ],
  "breadcrumb" : {
    "@id" : "https://prismalims.com/sample-holding-times#breadcrumb"
  },
  "dateModified" : "2026-11-26",
  "description" : "Sample holding times in plain words: when the clock starts, common EPA hold times with sources, prep and analysis clocks, and how a LIMS tracks them.",
  "inLanguage" : "en-US",
  "isPartOf" : {
    "@id" : "https://prismalims.com/#website"
  },
  "mainEntity" : {
    "@id" : "https://prismalims.com/sample-holding-times#article"
  },
  "name" : "Sample Holding Times: How a LIMS Tracks Them - Prisma LIMS",
  "url" : "https://prismalims.com/sample-holding-times"
}
```

```json
{
  "@context" : "https://schema.org",
  "@id" : "https://prismalims.com/sample-holding-times#article",
  "@type" : "Article",
  "author" : {
    "@id" : "https://prismalims.com/#organization",
    "@type" : "Organization",
    "name" : "Prisma LIMS"
  },
  "dateModified" : "2026-11-26",
  "datePublished" : "2026-11-24",
  "headline" : "Sample holding times: how a LIMS keeps the clock",
  "image" : "https://48211686.fs1.hubspotusercontent-na1.net/hubfs/48211686/website%20assets/images/featured%20images/og-sample-holding-times.jpg",
  "inLanguage" : "en-US",
  "mainEntityOfPage" : {
    "@id" : "https://prismalims.com/sample-holding-times#webpage"
  },
  "publisher" : {
    "@id" : "https://prismalims.com/#organization"
  }
}
```

```json
{
  "@context" : "https://schema.org",
  "@id" : "https://prismalims.com/sample-holding-times#breadcrumb",
  "@type" : "BreadcrumbList",
  "itemListElement" : [ {
    "@type" : "ListItem",
    "item" : "https://prismalims.com/",
    "name" : "Home",
    "position" : 1
  }, {
    "@type" : "ListItem",
    "item" : "https://prismalims.com/environmental-lims",
    "name" : "Environmental LIMS",
    "position" : 2
  }, {
    "@type" : "ListItem",
    "name" : "Sample holding times",
    "position" : 3
  } ]
}
```

```json
{
  "@context" : "https://schema.org",
  "@id" : "https://prismalims.com/sample-holding-times#terms",
  "@type" : "DefinedTermSet",
  "hasDefinedTerm" : [ {
    "@id" : "https://prismalims.com/sample-holding-times#term-holding-time",
    "@type" : "DefinedTerm",
    "alternateName" : "Hold time",
    "description" : "A sample holding time is the longest time allowed between collecting a sample and preparing or analyzing it.",
    "inDefinedTermSet" : {
      "@id" : "https://prismalims.com/sample-holding-times#terms"
    },
    "name" : "Sample holding time",
    "url" : "https://prismalims.com/sample-holding-times#what"
  } ],
  "name" : "Holding time terms"
}
```